Emory University
Radiation Safety Office
Laboratory Worker Training Manual
RADIATION PROTECTION
External Exposure
In addition to properly
wearing monitoring devices, it is also the responsibility of the radiation
worker to control their external exposure to radiation sources. This
can be accomplished by following the three principles of radiation protection.
Time
Distance Shielding
By minimizing the amount of time spent around a radiation source, exposures
can be minimized as shown by the graph of exposure vs time. Prior
planning is the key to minimizing time spent around radiation. Preplan
the procedure, practice the technique without the radiation source and prepare
the tools, equipment and chemicals prior to work. These will markedly
reduce exposure time.
The next figure shows how radiation drops off with increased distance.
The relationship between radiation and distance is an inverse square relationship.
If the distance from a point source of radiation is doubled, the exposure
is only 1/4 of that at the original distance. Exposure can be minimized
by taking this into consideration when choosing storage areas and work
spaces.
The final principle of radiation protection is shielding. Radiation
interacts with the material which it passes through so by properly choosing
material placed between the radiation worker and the radiation, the amount
of radiation can be reduced. In choosing shielding, one should consider
the type of radiation and the energy of the radiation. Lead is a
very dense material and provides efficient shielding for x-rays and gamma
radiation. High energy beta particles such as those found in 32P
should be shielded with a less dense material such as plexiglass.
Internal Protection
In addition to performing
the necessary bioassay, it is also important for radiation workers to
protect against internal exposure to radiation. Internal exposure
occurs when radioactive material enters the body by ingestion, inhalation
or absorption through the skin. Special precautions can be taken
to greatly reduce the possibility of internalizing radioactive material.
To prevent ingestion
of radioactive material no eating, drinking, smoking or application of
cosmetics is allowed in areas where radioactive materials are used.
No food can be stored in refrigerators where radioactive material is also
stored. Also, to protect against ingestion, pipetting by mouth is
prohibited.
Some radioactive material such as unbound 125I, 131I
and some 35S compounds are volatile; therefore, to prevent
inhalation, these should be used under the hood. During iodination
charcoal masks and personal air monitors should be worn by persons performing
the iodination. Technique is checked by use of thyroid bioassay
before and after the procedure.
To prevent against absorption
of radioactive material, proper use of protective clothing is necessary.
Lab coats, disposable gloves and protective eyewear should be worn at
all times while working with radioactive material. Protective clothing
should be removed and hands should be washed carefully before leaving
the laboratory.
Laboratory
Surveys and Instrumentation
One of the best methods
for protection against internal and external exposure is the lab survey.
If performed correctly, the survey can detect areas of contamination which
could be sources of exposure. For example, if the phone is accidentally
contaminated by someone who answers it wearing contaminated gloves, the
lab survey will find the contamination before it becomes a problem.
Every time radioactive material is used in the laboratory, it is good
practice to survey the area where the material was used as well as survey
the person who used it. This is easily accomplished if the radioactive
material is a high energy beta emitter such as 32P or a gamma
and x-ray emitter such as 125I. Upon completion of the
experiment, use the survey meter to check the areas that were used and
then survey yourself, the cuffs of your lab coat, your hands and the soles
of your shoes. This survey does not have to be documented each time
it is done. Weekly, if radioactive material is used, a more thorough
lab survey is required. Survey forms are available from the health
physicist which include a map of your area and spaces for the instruments
used and results from your survey. The survey consists of a dose
rate survey and a wipe test survey. The record must include the
date, area surveyed, equipment used, name and signature of the person
conducting the survey, a diagram of the areas surveyed and measurements.
To begin the survey,
if the isotope is detectable with a survey meter, check the areas of the
lab where work has been performed and where isotopes are stored.
It is also a good practice to check areas that are not supposed to contain
radiation, the phone, the door knob, desks, "cold" centrifuges and incubators,
etc. This part of the survey may surprise you. It is very
important to use the survey meter appropriate for your isotope use.
The action level for dose rate surveys is 2 mR/hr in unrestricted areas,
on personal clothing and on skin while the action level for restricted
areas and protective clothing used only in restricted areas is 10 mR/hr.
This means that if the GM survey yields levels above the action level,
you should decontaminate or shield the area and notify radiation safety.
The survey meter that is best known is the Geiger Counter or GM
Survey Meter. It is simple in principle, easy to operate, relatively
inexpensive, sensitive, reliable and versatile. It is particularly
suited to radiation safety surveys. The GM can detect gamma or x-ray
radiation (125I and 51Cr) and high energy betas
(32P, 90Sr). The GM will not detect the 3H
betas and a sensitive probe such as the pancake is necessary to detect
35S and 14C.
The GM tube consists
of a cylinder with a conductive shell filled with an inert gas at low
pressure (1/8 atmosphere). It has a central wire with a positive
charge. Anytime an ionization occurs, it initiates a chain reaction
in which there is a succession of ionizations that cause the control wire
to collect a multitude of electrons. Multiplication of charge produces
a signal of around one volt which is used to activate the counting circuit.
When preparing to use
your GM there are four things which need to be checked:
Calibration
Battery
Response to radiation
Scale
Before performing the survey, check that the meter is in calibration.
Survey meters must be calibrated annually. A sticker on the side
of the meter shows the date of calibration and the date that the meter
is due for calibration. If the meter is out of calibration, contact
the Radiation Safety Office to have it calibrated. If possible,
RSO will loan you a meter while yours is being calibrated. After
calibration, the meter is returned to you with a certificate of calibration.
This certificate should be kept in a file in the laboratory.
Then, a battery check should be performed. Most of the meters have
a scale labeled "BAT." By scaling your meter to this setting, the
needle should move to the corresponding area on the meter face.
Never use an instrument with low battery power since this may result in
erroneous readings. The batteries can be changed in the instrument
when necessary without disturbing the calibration.
Next, a check that the meter is responding to radiation is necessary.
By placing the probe next to a known radiation source such as a stock
vial, the meter can be checked for response. If the meter is not
responding to radiation, contact the Radiation Safety Office for repair.
Finally, ensure that you are familiar with reading the meter. Sometimes
this is the most difficult thing about operating the survey meter.
Most instruments have three or more scales which represent multipliers
to use with the instrument scale. The most common are X0.1, X1,
X10 and X100. The instrument can be set initially at either the
highest or lowest setting and then moved as necessary to enable reading.
In most cases in the lab, the X0.1 or X1 will be the most appropriate
scale. You read the result from the meter and then multiply by the
scale that the meter is set at. Here are a few examples.
Example 1:
The needle registers at 3.6 . When the scales
are taken into account the readings are:
0.36 mR/hr on the 0.1 scale (3.6x0.1)
3.6 mR/hr on the 1 scale (3.6x1)
and 36 on the 10 scale (3.6x10).
Example
2:
The needle registers at 0.8 (see figure XX). When
the scales are taken into account the readings are:
0.08 mR/hr on the 0.1 scale (0.8x0.1)
0.8 mR/hr on the 1 scale (0.8x1)
8 mR/hr on the 10 scale (0.8x10)
100 mR/ hr on the 100 scale (1x100)
This meter has a special scale for the 100 setting.
The needle registers 1. When the 100 scale is taken into account
the radiation field is measured to be 100 mR/hr. It will be
under extremely odd circumstances that the 100 scale becomes useful
to you.
It is also a good practice to keep the audio on when your survey meter
is in use. Consider the following scenario. While working
the 32P, you absently turn on the GM not noticing which scale
the meter is set at. You work, checking instruments and gloves for
contamination and the needle does not move. Then you notice that
the scale is set at 100 and that is the reason that the needle has not
moved. When you turn the audio on the meter begins to chirp repeatedly.
The audio responds to radiation in the same manner without regard to the
scale. No matter which scale you are set at, you know
a hot spot when you find it.
Detector probes used with survey
meters vary with the type of work being done and the isotopes in use.
The best probe for radiation safety surveys is the pancake
probe. This probe enables one to detect 14C and 35S
which is very difficult if not impossible with other probes. For
estimating exposure rate from a beta or low energy gamma or x-ray source,
an end window
probe would be suitable. Estimating exposure rate from a medium
to high energy gamma or x-ray source is best done with a side window probe.
If two isotopes of different types or energies are being used in the lab,
this probe can be used with the window open and closed to determine the
isotope involved in the even of contamination or a spill.
The final step in the
weekly laboratory survey is the wipe test. This part of the survey
will determine whether there is removable contamination in the lab.
Using some absorbent material, filter paper, cotton swabs, etc, smear
any area where radioactive material is used. Then, using the proper
detection device, count the wipe to determine if any isotope was removed.
Just as with the dose rate survey, it is a good practice to check areas
which should not contain any removable contamination, door knobs, refrigerator
doors, etc. The action levels for surface contamination are 200
dpm/100 cm2 in unrestricted areas and on personal clothing
and skin and 2000 dpm/100 cm2 in restricted areas and on protective
clothing used only in the restricted area. However, it is good practice
to decontaminate any area or piece of equipment above 200 dpm/100 cm2
anywhere in the lab. This can be accomplished using gloves
and some form of cleaning fluid.
It will be necessary
for you to determine which detection device is necessary for the isotope
that you are using. To detect beta emissions, the liquid scintillation
counter is the appropriate detector. The scintillation cocktail
converts the beta energy to light which is then multiplied through the
use of a photomultiplier tube. This is then measured to determine
the activity. In order to determine the activity of the sample counted,
it is necessary to determine the efficiency of the detector. Since
the detector cannot count every disintegration, the number of the counts
which the detector determines must be corrected using this efficiency.
To determine the efficiency of your counter, you will need to count a
sample of known activity and then count for background. By
knowing the activity of the standard and by using equation 4, you
can determine the efficiency.
4) Eff =
(cpm- bkg ) / dpm
In equation 4, cpm is the number
of counts per minute determined by the detector, bkg is the number of
background counts per minute and dpm is the activity of the standard in
disintegrations per minute. For example, in Radiation Safety we
use a H-3 standard to calculate efficiency. The known number of
counts determined by the detector is 272 cpm, the background counts is
10 cpm and the known activity of the standard is 414 dpm.
Eff = ( 272
- 10 ) / 414 = 63 %
Using equation 4, we determine
the efficiency of the detector to be 63%.
Finally, to determine
the activity of the wipe test, we solve equation 4 for activity (see equation
5).
5) dpm =
( cpm - bkg ) / Eff
Here, cpm is the number
of counts determined by the detector when the sample is counted.
For example, a wipe test is counted in the our liquid scintillation counter
and the number of counts is 1595 cpm, the background is 9 cpm and from
above, the efficiency is 63%. Using equation 5, the wipe is determined
to be 2504 dpm.
dpm = ( 1595
- 9 ) / 0.63 = 2517 dpm
Since the result exceeds the
contamination limit set for a restricted area, clean up is necessary.
Some liquid scintillation counters will perform the above calculations
for you. Our Packard counter will print out the results in dpm using
standards which are run daily. Consult the operating manual of your
counter for more information.
For counting wipe tests
in labs where gamma or x-ray emitters are used, it is necessary to use
a gamma counter. The gamma counter is most frequently a sodium iodide
crystal activated with thallium and optically coupled to a photomultiplier
tube. The thallium activator converts the energy absorbed in the
crystal to light. The efficiency calculation for the gamma counter
is the same as that of the liquid scintillation counter. Consult
the manuals which accompany your gamma counter for more information.
Back to Training
Manual | Previous | Next
Send comments to swilson@ehso.emory.edu
Copyright © Emory University.
All Rights Reserved.
manual design
by Stan Wilson
|